Coronary Artery Stent Design for Challenging Disease

Started
14th November 2011
Research Team
Georgios Ragkousis
Investigators
Neil Bressloff

A 28mm stent deployment in a patient specific right coronary artery- The expansion is projected to the angiographic image.

It is well known that, across all populations (based on geographic location, race, ethnicity, age, sex) coronary artery disease (CAD) is the most common cause of death. CAD is a result of atherosclerosis, and affects the coronary arteries that surround the heart. When the inner walls of the coronary arteries thicken due to a build-up of plaque (cholesterol, fatty deposits, calcium, etc.), the vessel narrows and blood flow through the vessel is reduced leading to less oxygen supply to the heart muscle. Percutaneous coronary intervention (PCI) or “stenting” has become the main method for treating coronary occlusions in recent years. Stents are cylindrical mesh shaped devices that undergo plastic deformation so as to scaffold reopened artery walls and restore physiologically healthy blood flow. PCI involves the intra-vascular insertion (typically, through the groin) and deployment of a stent on a balloon-catheter-guidewire system. Once deployed, however, the presence of the stent can lead to unwanted events including in-stent restenosis (ISR) and stent thrombosis (ST).

Although current generation stents have been shown to significantly reduce the rate of ISR, sub-optimal delivery of stents still occurs leading to ISR and, in particular, ST. This is especially a problem in challenging anatomy with long lesions. A common complication of PCI in such challenging cases is stent malapposition or incomplete stent apposition (ISA) which is the lack of contact between stent struts and the underlying wall. This procedural challenge is not new, yet there has been a paucity of data produced with the aim of improving the mechanical impact of the implanted stent platform within a challenging geometry.

A method has been setup to (i) reconstruct diseased patient specific coronary artery segments; (ii) run many simulations of the deployment of state of the art stents into these segments (the Finite Element Analysis software package Abaqus is used for modelling procedures and demands significant run-times and resources, however, using the 16 core nodes on Iridis, it is possible to perform the simulation of stent deployment through balloon inflation in approximately 60 hours, as opposed to the more than 7 days it would take on a desktop workstation) and (iii) assess the degree of stent malapposition. The aim now is to devise a stent delivery system that can mitigate this problem.

Categories

Life sciences simulation: Biomedical

Physical Systems and Engineering simulation: Biomechanics, Structural dynamics

Algorithms and computational methods: Finite elements, Optimisation

Simulation software: Abaqus

Programming languages and libraries: Python

Computational platforms: Iridis, Linux, Windows

Transdisciplinary tags: Design